CM14 Troubleshooting
Endress+Hauser 35
Diagnostic
code
Message text Description
S844 Process value
Measured value outside the specified range.
Possible reasons:
Sensor in air
Air cushion in the assembly
Incorrect sensor inflow
Sensor defective
Remedy:
Increase process value
Check measuring chain
Change sensor type
S910 Limit switch Limit switch energized
Diagnostic
code
Message text Description
M500 Not stable
Sensor calibration canceled.
Main measured value unsteady.
Possible reasons:
Sensor overaged
Sensor temporarily dry
Buffer value not constant
Remedy:
Check sensor, replace if required
Check buffer
M526
Cell const.
Sensor calibration warning.
Invalid cell constant, max. range reached.
Remedy:
Recalibrate
Enter cell constant according to factory specifications
Replace sensor
M528 Cell const.
Sensor calibration warning.
Min. possible cell constant undershot.
Remedy:
Recalibrate
Enter cell constant according to factory specifications
M914 USP alarm
USP alarm.
Conductivity limit value for USP exceeded.
Remedy:
Check process
M915 USP warning
USP warning.
Conductivity limit value for USP undershot.
Remedy:
Check process