6.2 Operation without operating menu
6.2.1 Position of operating elements
The operating keys and DIP switches are located on the electronic insert in the device.
A0023125
Fig. 8: HART electronic insert
1 Operating keys for lower range value (zero) and upper range value (span)
2 Green LED to indicate successful operation
3 Slot for optional local display
4+5 DIP switch only for Deltabar M
6 DIP switch for alarm current SW / Alarm Min (3.6 mA)
7 DIP switch for switching damping on/off
8 DIP switch for locking/unlocking parameters relevant to the measured value
Function of the DIP switches
on
off
SW / P2=High
SW /Alarm min
SW /
damping
1
2
345
on
off
Display
Zero
Span
HAR T
R
FIELD COMMUNICATION PROTOCOL
SW / P2=High
delta p only
SW /Alarm min
SW /
damping
8 765 4
3
1
2
Switches Symbol/
labeling
Switch position
"off" "on"
1 The device is unlocked.
Parameters relevant to the measured value
can be modified.
The device is locked.
Parameters relevant to the measured value
cannot be modified.