Levelflex FMP56, FMP57
26 Endress+Hauser
Installation from the side
Ø16
(Ø0.63)
Ø17...18
(Ø0.67...0.71)
Ø<26
(Ø<1.02)
mm(in)
A0014140
If installation from above is not possible, the Levelflex can also be mounted from the side.
In this case, always fix the rope probe (® ä 25) .
Support rod probe if the lateral loadbearing capacity is exceeded (® ä 21). Only fix rod probes at the probe
end (® ä 25).
Installation in underground tanks
A0014142
Use coax probe for nozzles with large diameters in order to avoid reflections at the nozzle wall.