Liquiline CM44P Mounting
Endress+Hauser 13
NOTICE
Incorrect mounting location in the cabinet, spacing regulations not observed
Possible malfunctions as a result of heat buildup and interference from neighboring devices!
Do not position the device directly above sources of heat. The temperature specification
must be observed.
The components are designed for convection-based cooling. Avoid heat buildup. Ensure
openings are not covered, e.g. by cables.
Observe the specified distances to other devices.
Physically separate the device from frequency converters and high-voltage devices.
Recommended installation direction: horizontal. The specified ambient conditions, and
particularly the ambient temperatures, only apply for horizontal installation.
Vertical orientation is also possible. However, this requires additional fixing clips at the
place of installation to hold the device in position on the DIN rail.
Recommended installation of power unit: to the left of the device.
The following minimum clearance specifications must be observed:
Distances at the side in relation to other devices incl. power units and to the wall of the
cabinet:
at least 20 mm (0.79 inch)
Distance above and below the device and depth distance (to control cabinet door or other
devices installed there):
at least 50 mm (1.97 inch)
min. 50 (1.97)
min. 20 (0.79)
min. 50 (1.97)
min. 20 (0.79)
min. 50 (1.97)
A0039736
 3 Minimum clearance in mm (in)