Commissioning Liquiline CM44P
46 Endress+Hauser
7.2.1 Setting the operating language
Configuring the language
If you have not already done so, close the housing cover and screw the device closed.
1. Switch on the supply voltage.
Wait for the initialization to finish.
2. Press the soft key: MENU.
3. Set your language in the top menu item.
The device can now be operated in your chosen language.
7.3 Basic setup
Making basic settings
1. Switch to the Setup/Basic setup menu.
Make the following settings.
2. Device tag: Give your device any name of your choice (max. 32 characters).
3. Set date: Correct the set date if necessary.
4. Set time: Correct the set time if necessary.
For quick commissioning, you can ignore the additional settings for outputs, relays
etc. You can make these settings later in the specific menus.
5. To return to the measuring mode: press the soft key for ESC for at least one second.
Your controller now works with your basic settings. The sensors connected use the
factory settings of the sensor type in question and the individual calibration settings
that were last saved.
If you wish to configure your most important input and output parameters in the Basic setup :
Configure the current outputs, relays, limit switches, controllers, device diagnostics and
cleaning cycles with the submenus which follow the time setting.