Mounting Liquiline CM44P
22 Endress+Hauser
16 Mounted display
NOTICE
Incorrect installation
Damage, e.g. to the cable, or malfunctions are possible!
‣
Lay cables in such a way that they do not get squashed e.g. when closing the cabinet door.
‣
Plug the display cable only into the RJ45 socket in the base module. Otherwise, the display
will not function.
4.4 Post-mounting check
1. After installation, check the transmitter for damage.
2. Check whether the transmitter is protected against precipitation and direct sunlight
(e.g. by the weather protection cover).
3. After mounting, check all devices (controller, power unit, display) for damage.
4. Verify that the specified installation clearances have been observed.
5. Verify that all securing clips have been snapped into place and that the components are
securely positioned on the DIN rail.
6. Ensure that the temperature limits are observed at the mounting location.