Electrical connection Liquiline CM44P
34 Endress+Hauser
98i97i88i87i
98i97i88i87i
BN
WH
GN
YE
Sensor
BN
WH
GN
YE
Sensor
2DS
Ex-i
A0045659
 35 Sensors without additional supply voltage at sensor communication module type 2DS Ex-i
Intrinsically safe sensors for use in explosive atmospheres may only be connected to the
sensor communication module type 2DS Ex-i. Only the sensors covered by the certificates
may be connected (see XA).
5.3 Connecting additional inputs, outputs or relays
L
WARNING
Module not covered
No shock protection. Danger of electric shock!
Change or extend the hardware for the non-hazardous area: always fill the slots from left
to right. Do not leave any gaps.
If all of the slots are not occupied in the case of devices for the non-hazardous area:
always insert a dummy cover or end cover in the slot to the right of the last module. This
ensures that the unit is shock-protected.
Always ensure shock protection is guaranteed particularly in the case of relay modules (2R,
4R, AOR).
Hardware for the hazardous area may not be modified. Only the Manufacturer's Service
Team may convert a certified device to another certified device version. This includes all
modules of the transmitter with an integrated 2DS Ex-i module, as well as changes that
concern non-intrinsically safe modules.
The terminal strip (cabinet unit) is used to connect the cable shields.
If additional shields are required, connect them with PE centrally in the control cabinet via
terminal blocks supplied by the customer.