XA00182F-F Liquiphant M, Liquiphant S FTL50(H)/51(H)/51C, FTL70/71
26 Endress+Hauser
Device type FTL70, FTL71, Basic specification, Position 11 (Application) = L, N, Y
Temperature
class
Basic specification, Position 7
(Electronics; Output)
Ambient temperature T
a
(ambient): electronics
Process temperature T
p
(process): sensor
T6 4 –50 to +40 °C –60 to +85 °C
A, 1, 2, 5, 6, 7, 8, 9 –50 to +60 °C
T5 4
1)
–50 to +55 °C –60 to +100 °C
A, 1, 2, 5, 6, 7, 8, 9
1)
–50 to +70 °C
T4 all
1)
–50 to +70 °C –60 to +135 °C
T3 all
1)
–50 to +70 °C –60 to +200 °C
T2 all
1)
–50 to +70 °C –60 to +230 °C (L)
all
1)
–50 to +70 °C –60 to +280 °C (N)
all
1)
–50 to +70 °C –60 to +300 °C (Y)
1) For restrictions → 8, 27
Device type FTL50(H)/51(H),
Basic specification, Position 5, 6 (Probe Length; Type) = Ix, Qx, Jx, Kx, Lx, Rx, Sx, Tx
Device type FTL51C,
Basic specification, Position 5, 6 (Probe Length; Type) = xK, xL, xM, xN, xS
T
a
T
p
T
a
T
p
40
60
0
90 120
–50 150
70
0
4 A32
1
5
–50
A0025549
7
T
a
Ambient temperature in °C
T
p
Process temperature in °C
A Additional temperature range for sensors with Basic specification, Position 5, 6 (Probe Length; Type)
FTL50(H) = Ix, Qx and FTL51(H) = Jx, Kx, Lx, Rx, Sx, Tx
1 Temperature separator or pressure tight feed through
2 Basic specification, Position 7 (Electronics; Output) = 4: –50 °C ≤ T
a
≤ +40 °C (T6)
3 Basic specification, Position 7 (Electronics; Output) = A, D: –50 °C ≤ T
a
≤ +60 °C (T6)
4 ECTFE (xK)
5 PFA, Enamel (xL, xM, xN, xS)