Liquistation CSF48
Endress+Hauser 11
Sampling with a flow
assembly
A flow assembly is integrated in the stand for sampling purposes.
The flow assembly is used when sampling in pressurized systems, e.g.:
• Containers located at a higher level
• Pressurized pipes
• Pumping with external pumps
The flow rate should be 1000 to 1500 l/h.
The outlet of the flow assembly must be unpressurized (e.g. drain, open channel)!
Use the ball valve 1 to set the flow rate to 1000 l/h to 1500 l/h. When the sampling cycle begins, you
can use one of the relay outputs to control and open valve 2. The medium flows through the pipe and
the flow assembly into the outflow. When an adjustable delay time has elapsed, the sample is taken
directly from the flow assembly. Valve 2 is closed again once the sample has been taken.
Valve 1 and 2 are not included in the scope of delivery
(order code: 71180379).
a0013127
Flow assembly
1
2
3
Flow assembly inlet: ¾"
Sample connection
Flow assembly outlet: 1¼"
(can also be ordered separately as kit no.: 71119408)
a0013458
Example: Sampling from pressure piping
1 Ball valve 1
2Valve 2
3 Flow assembly integrated into the stand