Installation conditions for
Samplefit CSA420 sampling
assembly
Note the following when installing the sampling assembly in a pipe:
The best installation location is in the ascending pipe (pos. 2). Installation is also possible in the
horizontal pipe (pos. 1).
Avoid installation in the down pipe (pos. 4).
Avoid siphoning effects in the suction line.
The minimum vertical distance between the assembly and the inlet of the sampler must be at least
0.5 m (1.65 ft).
Note the following when erecting the sampler:
Erect the device on a level surface.
Protect the device against additional heating (e.g. from heaters).
Protect the device against mechanical vibrations.
Protect the device against strong magnetic fields.
Make sure air can circulate freely at the side panels of the cabinet. Do not mount the device directly
against a wall. Allow at least 150 mm (5.9") from the wall to the left and right.
Do not erect the device directly above the inlet channel of a wastewater treatment plant.
a0015886
Installation conditions for Liquistation CSF48 with Samplefit CSA420 sampling assembly