Installation conditions for
Samplefit CSA420 sampling
assembly
Note the following when installing the sampling assembly in a pipe:
• The best installation location is in the ascending pipe (pos. 2). Installation is also possible in the
horizontal pipe (pos. 1).
• Avoid installation in the down pipe (pos. 4).
• Avoid siphoning effects in the suction line.
• The minimum vertical distance between the assembly and the inlet of the sampler must be at least
0.5 m (1.65 ft).
Note the following when erecting the sampler:
• Erect the device on a level surface.
• Protect the device against additional heating (e.g. from heaters).
• Protect the device against mechanical vibrations.
• Protect the device against strong magnetic fields.
• Make sure air can circulate freely at the side panels of the cabinet. Do not mount the device directly
against a wall. Allow at least 150 mm (5.9") from the wall to the left and right.
• Do not erect the device directly above the inlet channel of a wastewater treatment plant.
a0015886
Installation conditions for Liquistation CSF48 with Samplefit CSA420 sampling assembly