Liquistation CSF48
30 Endress+Hauser
Mounting conditions for
Liquistation CSF48
1. Correct
The suction line must be routed with a downward gradient to the sampling point.
2. Incorrect
The sampler should never be mounted in a place where it is exposed to aggressive gases.
3. Incorrect
Avoid siphoning effects in the suction line.
4. Incorrect
The suction pipe should never be routed with an upward gradient to the sampling point.
Note the following when erecting the device:
• Erect the device on a level surface.
• Protect the device against additional heating (e.g. from heaters).
• Protect the device against mechanical vibrations.
• Protect the device against strong magnetic fields.
• Make sure air can circulate freely at the side panels of the cabinet. Do not mount the device directly
against a wall. Allow at least 150 mm (5.9") from the wall to the left and right.
• Do not mount the device directly above the inlet channel of a wastewater treatment plant.
a0011693
Mounting conditions for Liquistation CSF48 for open channels