Liquistation CSF48
26 Endress+Hauser
EtherNet/IP
Web server The web server allows complete access to the device configuration, measured values, diagnostic
measurements, logbooks and service data via standard WiFi/WLAN/LAN/GSM or 3G router with a
user-defined IP address.
Power supply
Electrical connection --> For a detailed wiring diagram, see the Operating Instructions for Liquistation CSF48
Supply voltage Depending on version:
• 100 to 120/200 to 240 V AC ±10 %, 50/60 Hz
• 24 V DC +15/-9 %
Broadcast support for function codes 06, 16, 23
Output data 16 measured values (value, unit, status), 8 digital values (value,
status)
Input data 4 set points (value, unit, status), 6 digital values (value, status),
diagnostic information
Supported features Address can be configured using DHCP or software
Protocol EtherNet/IP
ODVA certification Yes
Device profile Generic device (product type: 0x2B)
Manufacturer ID 0x049E
h
Device type ID 0x109C
h
Polarity Auto-MIDI-X
Connections (max.) CIP 12
I/O 6
Explicit
message
6
Multicast 3 consumers
Minimum RPI 100 ms (default)
Maximum RPI 10000 ms
System integration EtherNet/IP EDS
Rockwell Add-on-Profile Level 3, Faceplate for
FactoryTalk SE
IO data Input
(T --> O)
Device status and diagnostic message with
highest priority
Measured values:
• 16 AI (analog input) + Status + Unit
• 8 DI (discrete input) + Status
Output
(O --> T)
Actuating values:
• 4 A0 (analog output) + status + unit
• 8 DO (discrete output) + Status
TCP port 80
Supported features • Remote-controlled device configuration
• Save/restore device configuration
• Logbook export (file formats: CSV, FDM)
• Access to web server via DTM or Internet Explorer