The device does not have a power switch
A fuse with a maximum rating of 10 A must be provided by the customer. Observe the local
regulations for installation.
For samplers with CSA approval, use a 10 A 250 V AC HBC fuse.
Cable entry Depending on version:
1 x M25, 7 x M20 cable gland
1 x M25, 1 x M20 cable gland
Permitted cable diameter:
M20x1.5 mm: 7 to 13 mm (0.28 to 0.51")
M25x1.5 mm: 9 to 17 mm (0.20 to 0.67")
Mains fuse Optional fuses on DIN rail
T10A (for 24V power supply)
T3.15A (for 230V power supply)
T10A (fuse for battery backup)
For version with cCSAus approval: T4A (for temperature regulator)
Power consumption Version with vacuum pump: 290 VA
Version with peristaltic pump: 290 VA
Version with sampling assembly: 290 VA
Version with 24V power supply: 240 W
Power failure Power supply via battery (optional): 2 x 12 V, 7.2 Ah, with additional charge controller
Replace batteries with battery type Panasonic LC-R127R2PG1.
Real-time clock: lithium battery, type CR2032
Performance characteristics
Sampling methods Vacuum pump / peristaltic pump / sampling assembly:
Event sampling
Single and multiple samples
•Sampling table
Vacuum pump / sampling assembly:
In proportion to time
In proportion to volume
Peristaltic pump:
In proportion to time
In proportion to volume
In proportion to flow
Dosing volume Vacuum pump:
20 to 350 ml (0.7 to 12 fl.oz.)
Peristaltic pump:
10 to 10,000 ml (0.3 to 340 fl.oz.)
A sample volume < 20 ml can vary in dosing accuracy and repeatability, depending on the
application.
Sampling assembly:
10, 30 or 50 ml (0.3; 1 or 1.7 fl.oz.)
Dosing accuracy Vacuum pump:
±5 ml (0.17 fl.oz.) or 5 % of the set volume
Peristaltic pump:
±5 ml (0.17 fl.oz.) or 5 % of the set volume
Sampling assembly:
±2 ml (0.07 fl.oz.) or 5 % of the set volume