Electrical connection Liquisys M CLM223/253
16 Endress+Hauser
4 Electrical connection
WARNING
!
Device is energized
Improper connection can cause injury or death.
The electrical connection must only be carried out by a certified electrician.
Technical personnel must have read and understood the instructions in this manual and
must adhere to them.
Prior to beginning any wiring work, make sure voltage is not applied to any of the cables.
4.1 Wiring
4.1.1 Connecting supply voltage
NOTICE
The device does not have a power switch
You must provide a protected circuit breaker in the vicinity of the device.
This must be a switch or a power-circuit breaker and you must label it as the circuit
breaker for the device.
At the supply point, the power supply for the 24 V versions must be isolated from
dangerous live cables by double or reinforced insulation.