Commissioning Liquisys M CLM223/253
32 Endress+Hauser
6.3 Quick start-up
After power-up, you must make some settings to configure the most important functions of
the transmitter which are required for correct measurement. The following section gives an
example of this.
User input Setting range
(Factory settings,
bold)
Display
1. Press the
F
key.
2. Enter the code 22 to edit the setup. Press
F
.
3. Press
S
until you get to the "Service" function
group.
4. Press
F
to be able to make your settings.
5. In S1, select your language, e.g. "ENG" for English.
Press
F
to confirm.
ENG = English
GER = German
FRA = French
ITA = Italian
NEL = Dutch
ESP = Spanish
6. Press
P
simultaneously to exit the "Service"
function group.
7. Press
S
until you get to the "Setup 1" function
group.
8. Press
F
to be able to make your settings for
"Setup 1".
9. In A1, select the desired mode of operation, e.g.
"cond" = conductive.
Press
F
to confirm.
cond = conductive
ind = inductive
MOhm = resistivity
conc = concentration
10. In A2, press
F
to confirm the factory settings.
(if A1 = conc, else step 12)
%
ppm
mg/l
TDS = Total Dissolved
Solids
none
11. In A3, press
F
to confirm the factory settings.
XX.xx
X.xxx
XXX.x
XXXX
12. In A4, press
F
to confirm the factory settings.
auto, μS/cm, mS/cm,
S/cm, μS/m, mS/m,
S/m
13. In A5, enter the cell constant of the connected
sensor.
Refer to the sensor’s quality certificate for the
exact value.
cond: 1.000 cm
-1
ind: 1.98 cm
-1
MOhm: 0.01 cm
-1
0.0025 to 99.99 cm
-1
S
SERVICE
S1
Language
ENG
A
SETUP 1
A1
Oper.Mode
cond
A2
Conc.Unit
%
A3
Format
XX.xx
A4
Unit
auto
A5
1/cm
Cellconst
1.000