Diagnostics and troubleshooting Liquisys M CLM223/253
72 Endress+Hauser
7 Diagnostics and troubleshooting
7.1 Troubleshooting instructions
The transmitter constantly monitors its functions itself. If an error occurs which the device
recognizes, this is indicated on the display. The error number is shown below the display of
the main measured value. If more than one error occurs, you can call these up with the
MINUS key.
Refer to the "System error messages" table for the possible error numbers and remedial
measures.
Should a malfunction occur without any transmitter error message, please refer to the
"Process-specific errors" or the "Device-specific errors" tables to localize and rectify the error.
These tables provide you with additional information on any spare parts required.
7.2 System error messages
The system error messages can be called up and selected with the MINUS key.
Error
no.
Display Tests and/or remedial measures Alarm contact Error current Autom.
cleaning start
PROFIBUS
status
Facty User Facty User Facty User PV
1)
Temp
E001 EEPROM memory error 1. Switch device off and then on again.
2. Load device software compatible
with the hardware (with optoscope,
see "Optoscope service tool" section).
3. Load measurement-parameter
specific device software.
4. If the error persists, send in the
device for repair to your local
service organisation or replace the
device.
Yes No
2)
0C 0C
E002
Instrument not calibrated,
calibration data invalid, no
user data, user data
invalid (EEPROM error),
instrument software not
suitable to hardware
(controller)
Yes No
2
0C 0C
E003 Download error Invalid configuration. Repeat download. Yes No No 0C 0C
E004
Instrument software
version not compatible
with module hardware
version
Load software compatible with
hardware.
Load measurement-parameter specific
device software.
Yes No No 0C 0C
E007
Transmitter malfunction,
instrument software not
compatible with
transmitter version
Yes No
2
0C 0C
E008
Sensor or sensor
connection faulty
Check sensor and sensor connection
(Service).
Yes Yes No 0C 0C
E010
Temperature sensor
defective, not connected
or short-circuited
(measurement continues
with 25 °C)
Check temperature sensor and
connections; check device and
measuring cable with temperature
simulator if necessary.
Check correct option selected in field
B1.
Yes No No 80 0C
E025
Limit for Airset offset
exceeded
Repeat Airset (in air) or replace sensor.
Dry sensor.
No No No 80 80
E036
Calibration range of
sensor exceeded
Clean sensor and recalibrate; if
necessary, check sensor and
connections.
No No No 80 80
E037
Below calibration range of
sensor
No No No 80 80
E045 Calibration aborted Recalibrate No No
2
80 80