Liquisys M CLM223/253 Commissioning
Endress+Hauser 71
C126
Store calibration
results?
yes
no
new
If C125 = E xxx, then only no or new.
If new, return to C.
If yes/no, return to “Measurement”.
C1 (3)
Calibration with
sensor adaptation for
inductive sensors
(Plus Package only)
Airs = Airset (1)
Cellc = cell constant (2)
InstF = installation
factor (3)
Sensor calibration with compensation of
wall influence.
On inductive sensors, the distance from
the sensor to the wall of the pipe and the
material of the pipe (conductive or
nonconductive) influence the measured
value. The installation factor shows this
influence. See the technical information
of the installed sensor.
The sensor is installed in the process.
C131
Enter process
temperature (MTC)
25 °C
-35.0 to 250.0 °C
Only exists if B1 = fixed.
C132
Enter
value of the
calibration solution
2.10 %/K
0.00 to 20.00 %/K
This value is specified in the TI of all E+H
calibration solutions. You can also use the
printed-on table to calculate the value.
Set to 0 for calibration with
uncompensated values.
C133
Enter correct
conductivity value of
the calibration
solution
current measured value
0.0 μS/cm to 9999 mS/cm
You should select a value close to the
application range.
C134
Calculated
installation factor is
displayed
1
0.10 to 5.00
On inductive sensors, the distance from
the sensor to the wall of the pipe and the
material of the pipe (conductive or non-
conductive) influence the measured
value. The installation factor shows this
influence. See the technical information
of the installed sensor.
C135
Calibration status is
displayed
o.k.
E xxx
If the calibration status is not o.k., the
second display line shows an explanation
of the error.
C136
Store calibration
results?
yes
no
new
If C135 = E xxx, then only no or new.
If new, return to C.
If yes/no, return to “Measurement”.
Coding Field Selection or range
(factory settings bold)
Display Info