Liquisys M CLM223/253 Commissioning
Endress+Hauser 33
14. In A6, enter the resistance of the cable
(conductive sensors only).
0
0 to 99.99
15. If you need to stabilize the display, enter the
required damping factor in A7.
Press
F
to confirm.
The display returns to the initial display of "Setup
1"
1
1 to 60
16. Press
S
to go to the "Setup 2" function group
17. Press
F
to edit "Setup 2"
18. In B1, select your temperature sensor.
Press
F
to confirm.
Pt 100
Pt 1k = Pt 1000
NTC30
fixed
19. In B2, select the appropriate temperature
compensation for your process, e.g. "lin" = linear.
Press
F
to confirm your selection.
For detailed information on temperature
compensation, see chapter "Setup 2"
none
lin = linear
NaCl = common salt
(IEC 60746)
Pure = ultra pure
water NaCl
PureH = ultra pure
water HCl
Tab = table
20. In B3, enter the temperature coefficient .
Press
F
to confirm.
2.1 %/K
0.0 ... 20.0 %/K
21. The real temperature is displayed in B5. If
necessary, calibrate the temperature sensor to an
external measurement.
Press
F
to confirm.
Display and entry of
real temperature
-35.0 to 250.0 °C
fixed
22. The difference between the measured and the
entered temperatures is displayed.
Press
F
.
The display returns to the initial display of the
"Setup 2" function group.
0.0 °C
-5.0 to 5.0 °C
23. Press
S
to go to the "Current output" function
group.
24. Press
F
to edit the output settings.
User input Setting range
(Factory settings,
bold)
Display