Commissioning Liquisys M CLM223/253
34 Endress+Hauser
For inductive sensors you must perform an airset before installation of the sensor, refer
to the chapter "Calibration"
25. In O1, select your output, e.g. "Out1" = output 1.
Press
F
to confirm.
Out1
Out2
26. In O3, select the linear characteristic.
Press
F
to confirm.
lin = linear (1)
sim = simulation
Tab = table
27. In O311, select the current range for your output,
e.g. 4 to 20 mA.
Press
F
to confirm.
4 to 20 mA
0 to 20 mA
28. In O312, enter the conductivity corresponding to
the minimum current value at the transmitter
output, e.g. 0 μS/cm.
Press
F
to confirm.
cond/ind: 0.00 μS/cm
MOhm: 0.00 kcm
Conc: 0.00 %
Temp: 0.00 °C
29. In O313, enter the conductivity corresponding to
the maximum current value at the transmitter
output, e.g. 2000 mS/cm.
Press
F
to confirm.
The display returns to the initial display of the
"Current output" function group.
cond/ind: 2000 mS/
cm
MOhm: 500 kcm
Conc: 99.99 %
Temp: 150 °C
30. Press
P
simultaneously to switch to the
measurement mode.