Electrical connection Liquisys M CLM223/253
18 Endress+Hauser
Pay attention to the following:
The device is approved for protection class II and is generally operated without a protective
earth connection.
To guarantee measuring stability and functional safety, you have to ground the outer
screen of the sensor cable:
Inductive sensors: terminal "S"
Conductive sensors: PE distributor rail
This is on the cover frame for panel-mounted instruments and in the connection
compartment for field devices. Ground the PE distributor rail or the ground terminal.
The circuits "E" and "H" are not galvanically isolated from each other.
Field instrument connection
Guide the measuring cables through the PG glands into the housing. Connect the measuring
cables in accordance with the terminal assignment ( å 10).
a0008915
Fig. 10: Field instrument connection compartment sticker
1 PE distributor rail for CD/CS version (conductive sensors)
42
+++++-----
434154 55 56
NC
12
13
11
31
83 84 15
32
16
3433
S
NC 938685 94 81 82 NC
51 52 53 57 58 59 47 48 49
Lf
Temp.
(opt.)
+15V
10mA
WH
YE
GN
REL 4SENSOR REL 3 REL 2 REL 1
131095-4F
2
1
11 12
13
NC
Mains
Hilfsenergie
L1
+
N
-
(AC)
(DC)
Contacts:
max. 2A
AC: 250V/500VA
DC: 30V/60W
+-
23 24
4..20mA
PE
1
Binär/
Binary
Binär/
Binary