Electrical connection Liquisys M CLM223/253
20 Endress+Hauser
Structure and termination of the measuring cables
a0008938-de
Fig. 12: Structure of the special measuring cables
A Semiconductor layer
sc Screening
For further information on the cables and junction boxes, please refer to the
"Accessories" section.
Field instrument measuring cable connection
Proceed as follows to connect a conductivity sensor to the field instrument:
1. Open the housing cover to access the terminal block in the connection compartment.
2. Break the punching of a cable gland from the housing, mount a cable gland and guide
the cable through this cable gland.
3. Connect the cable in accordance with the terminal assignment (see connection
compartment sticker).
4. Tighten the cable gland.