Liquisys M CLM223/253 Commissioning
Endress+Hauser 43
The values for the sample characteristic ( å 27) are entered in the following table. The
distance per mA can be calculated from signal / mA.
First enter the desired current output configuration into the following blank table with a
pencil. Calculate the resulting signal distance per mA to observe the necessary minimum
slope. Then enter the values in the device.
Current output 1 Current output 2
Value pair [mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
15004
2 1000 16 41.66
3 2000 20 250
Current output 1 Current output 2
Value pair [mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
[mS/cm]
[%] [°C]
Current
[mA]
Distance
per mA
1
2
3
4
5
6
7
8
9