Liquisys M CLM223/253 Maintenance
Endress+Hauser 83
Conductivity
For conductivity, the values in the following table are valid, if the cell constant k is set to the
nominal value according to column 2.
Otherwise: Display conductivity [mS/cm] = k[cm
-1
] 1 / R[k]
The M measurement is normally used for pure and ultrapure water and therefore is
only wise for cell constants where k= 0.01 or k = 0.1 cm
-1
.three-line system.
Resistance R Cell constant k Display for conductivity Display for M
10
1 cm
-1
100 mS/cm
10 cm
-1
1000 mS/cm
100
0.1 cm
-1
1 mS/cm 1 k cm
1 cm
-1
10 mS/cm
10 cm
-1
100 mS/cm
1000
0.1 cm
-1
0.1 mS/cm 10 k cm
1 cm
-1
1 mS/cm
10 cm
-1
10 mS/cm
10 k
0.01 cm
-1
1 μS/cm 1 M cm
0.1 cm
-1
10 μS/cm 100 k cm
1 cm
-1
100 μS/cm
10 cm
-1
1 mS/cm
100 k
0.01 cm
-1
0.1 μS/cm 10 M cm
0.1 cm
-1
1 μS/cm 1 M cm
1 cm
-1
10 μS/cm
1 M
0.01 cm
-1
0.01 μS/cm 100 M cm
0.1 cm
-1
0.1 μS/cm 10 M cm
1 cm
-1
1 μS/cm
10 M
0.01 cm
-1
0.001 μS/cm
0.1 cm
-1
0.01 μS/cm 100 M cm