Commissioning Liquisys M CUM223/253
32 Endress+Hauser
14. In B1, switch wiper controller on or off.
Press ENTER to confirm.
off
on
auto
15. In B2, enter operation period of wiper.
Press ENTER to confirm.
30 s
3 to 999 s
16. In B3, enter pause time between two wiping
cycles.
Press ENTER to confirm.
30 min
1 to 7200 min
17. In B4, select the calibration data set to be used.
Press ENTER to confirm.
3
1 to 3
18. In B5, select the data set to be copied.
Press ENTER to confirm.
no
1 -> 2
1 -> 3
2 -> 3
3 -> 2
19. In B6, select if the measured value shall be
displayed with reflection adjustment (for
solutions 2 FNU / 5 ppm).
Press ENTER to confirm.
yes
no
20. In B7, enter the corrected temperature of the
temperature sensor (for adjustment to external
measurement).
Press ENTER to confirm.
Current measured
value
-5.0 to 100.0 °C
21. In B8, the current offset is displayed.
Press ENTER to confirm.
Current offset
-5.0 to 5.0 °C
22. In B9, enter gas bubble barrier. For clear media
with measured values <1000 NTU the gas bubble
barrier has to be set to 100 %.
Press ENTER to confirm.
3.0 %
0.1 to 100 %