Commissioning Liquisys M CUM223/253
34 Endress+Hauser
6.4 System configuration
6.4.1 Setup 1 (Turbidity)
In the SETUP 1 function group, you can change the operating mode and the sensor settings.
Coding Field Selection or range
(factory settings bold)
Display Info
A
Function group
SETUP 1
Basic settings.
A1
Select operating
mode
FNU
NTU
ppm
mg/l
%
spec.
Any change in operating mode causes an
automatic reset of user defined settings.
The offsets for turbidity and temperature
are reset to zero.
A2 Select display unit
kg/l
%
t/m
A2 is only accessible, if A1 = spec.
A3 Select display format
XX.xx
X.xxx
XXX.x
XXXX
A3 is only accessible, if A1 = spec.
A4
Connected sensor is
displayed
CUS31
CUS41
The transmitter automatically detects
which sensor is connected.
A5
Enter measured
value damping
10
1 to 60
Measured value damping causes averaging
over the specified number of individual
measured values. It is used, for example, to
stabilize the display with applications that
fluctuate a great deal.
There is no damping if “1” is entered.
A
SETUP 1
A1
FNU
Oper.Mode
A2
kg/l
Conc.Unit
A3
XX.xx
Format
A4
CUS31
sensor
A5
10
Damping