Commissioning Liquisys M CUM223/253
74 Endress+Hauser
6.8 Slope
With the settings in the function group SLOPE, a measured value can be adapted to a
reference value. All measured values are proportionally adapted over the entire measuring
range according to this change.
Example:
The displayed measured value is 2.5 g/l. It is adapted to the reference value of 2.0 g/l by
means of the slope function. The change is 20 %, i.e. all measured values are reduced by
20 % over the entire measuring range.
An OFFSET having been edited before is reset to the factory setting. In contrast to the
one-point calibration, the edited slope can be reset by setting the slope factor to 1.0.
Coding Field Setting range
(Factory settings, bold)
Display Info
N
Function group
SLOPE
N1 Enter absolute value Current measured value
N2 Enter slope Current slope Slope is displayed, can be edited.
N3 Status is displayed
o.k.
Exxx
N4 Store slope?
yes
no
new
N
SLOPE
N1
0.000
Real PV
N2
1.000
PV Slope
N3
o.k.
Status
N4
yes
Store