Liquisys M CUM223/253 Diagnostics and troubleshooting
Endress+Hauser 75
7 Diagnostics and troubleshooting
7.1 Troubleshooting instructions
The transmitter constantly monitors its functions itself. If an error occurs which the device
recognizes, this is indicated on the display. The error number is shown below the display of
the main measured value. If more than one error occurs, you can call these up with the
MINUS key.
Refer to the "System error messages" table for the possible error numbers and remedial
measures.
Should a malfunction occur without any transmitter error message, please refer to the
"Process-specific errors" or the "Device-specific errors" tables to localize and rectify the error.
These tables provide you with additional information on any spare parts required.
7.2 System error messages
The system error messages can be called up and selected with the MINUS key.
Error
no.
Display Tests and/or remedial measures Alarm contact Error current Autom.
cleaning start
PROFIBUS
status
Facty User Facty User Facty User PV
1)
Temp
E001 EEPROM memory error 1. Switch device off and then on again.
2. Load device software compatible
with the hardware (with optoscope,
see "Optoscope service tool" section).
3. Load measurement-parameter
specific device software.
4. If the error persists, send in the
device for repair to your local
service organization or replace the
device.
Yes No — —
2)
0C 0C
E002
Instrument not calibrated,
calibration data invalid, no
user data, user data
invalid (EEPROM error),
instrument software not
suitable to hardware
(controller)
Yes No — —
2
0C 0C
E003 Download error
Invalid configuration. Repeat download,
check optoscope.
Yes No No 0C 0C
E004
Instrument software
version not compatible
with module hardware
version
Load software compatible with
hardware.
Load measurement-parameter specific
device software.
Yes No No 0C 0C
E007
Transmitter malfunction,
instrument software not
compatible with
transmitter version
Contact E+H Service. Yes No — —
2
0C 0C
E008
Sensor or sensor
connection faulty
Check sensor and sensor connection
(Service).
Check initialization of wiper.
Yes Yes No 0C 0C
E026 Wiper error
Check wiper and test function using
manual control if necessary.
Yes No No 44 80
E045 Calibration aborted Repeat calibration No No — —
2
80 80
E055
Below main parameter
measuring range
Check measurement and connections;
check device and measuring cable.
Yes No No 44 80
E057
Main parameter
measuring range exceeded
Yes No No 44 80
E059
Below temperature
measuring range
Yes No No 80 44
E061
Temperature measuring
range exceeded
Yes No No 80 44