Liquisys M CUM223/253 Commissioning
Endress+Hauser 67
Version TB: Initial settings for residual concrete water
The version TB contains the Plus-package with additional setting options. The default values
set in individual fields of the menus differ from the standard version to make commissioning
as simple as possible.
These values are so selected that no additional settings need to be made for applications in
residual concrete water. If you ever reset the device to the original factory settings (set
default), you can find valid values for residual concrete water in the table below.
Calibration is carried out in the % measuring range (the transmitter switches
automatically). With the above setting, a density of e.g. 1.12 kg/l = 12 % must be
selected. In some cases, it may be necessary to adapt the calibration to a real sample. To
do so, perform a single-point calibration.
Menu field Setting
Mode of operation A1 spec.
Unit A2 kg/l
Display format A3 XX.xx
Measured value damping A5 10
Calibration data record B4 3
Current output O1 Out1
Characteristic O2 lin
Current range O211 0 to 20 mA
Measured value 0 mA O212 1.00
Measured value 20 mA O213 1.30
Active characteristic K1 1
Edited characteristic K2 1
Table option K3 edit
Number of value pairs K4 2
Support point K5 1 to 2
Measured value / display value K6 / K7 1: 0%/1.00 2: 50%/1.50
Language version S1 GER