EasyManua.ls Logo

Endress+Hauser Micropilot FMR20 - Electromagnetic Compatibility (EMC)

Endress+Hauser Micropilot FMR20
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR20 HART
Endress+Hauser 21
Process
Process temperature, process
pressure
p
-40
+80 ( 76)+1
0 ( )+32
T
p
[ °F ]°C ( )
-1( )-14.5
0
[bar ](psi)
3 (43.5)
A0029007-EN
 22 FMR20: Permitted range for process temperature and process pressure
Process temperature range
–40 to +80 °C (–40 to +176 °F)
Process pressure range, threaded process connection
p
rel
= –1 to 3 bar (–14.5 to 43.5 psi)
p
abs
< 4 bar (58 psi)
Process pressure range, UNI flange process connection
p
rel
= –1 to 1 bar (–14.5 to 14.5 psi)
p
abs
< 2 bar (29 psi)
The pressure range may be further restricted in the event of a CRN approval.
Dielectric constant
For liquids
ε
r
≥ 4
Contact Endress+Hauser for lower ε
r
values
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
the Endress+Hauser DC manual (CP01076F)
the Endress+Hauser "DC Values App" (available for Android and iOS)

Table of Contents