Operating menu Micropilot FMR51, FMR52 FOUNDATION Fieldbus
126 Endress+Hauser
17.3 "Setup" menu
• : Marks the navigation path to the parameter via the display and operating
module.
• : Marks the navigation path to the parameter via an operating tool (e.g.
FieldCare).
• : Marks parameters which can be locked via the software locking → 48.
Navigation Setup
Distance unit
Navigation Setup → Distance unit
Description Select distance unit.
Selection SI units
• mm
• m
US units
• ft
• in
Factory setting m
Tank type
Navigation Setup → Tank type
Prerequisite Medium type (→ 138) = Liquid
Description Select tank type.
Selection • Bypass / pipe
• Stilling well
• Workbench test
• Open channel
• Sphere
• Storage vessel
• Process vessel standard
• Process vessel with agitator
• Wave guide antenna
Factory setting Depending on the antenna
Additional information Depending on the antenna some of the options mentioned above may not be available or
there may be additional options.