Optional specifications
Safety instructions:
General
Staff must meet the following conditions for mounting, electrical installation, commissioning and
maintenance of the device:
Be suitably qualified for their role and the tasks they perform
Be trained in explosion protection
Be familiar with national regulations (e.g. IEC/EN 60079-14)
Install the device according to the manufacturer's instructions and national regulations.
Do not operate the device outside the specified electrical, thermal and mechanical parameters.
Only use the device in media to which the wetted materials have sufficient durability.
Refer to the temperature tables for the relationship between the permitted ambient temperature for the
sensor and/or transmitter, depending on the range of application, and the temperature class.
Modifications to the device can affect the explosion protection and must be carried out by staff authorized
to perform such work by Endress+Hauser.
Safety instructions:
Special conditions
Permitted ambient temperature range at the electronics housing: –40 °C T
a
+80 °C.
Observe the information in the temperature tables.
Device type FMR50, FMR52, FMR53, FMR54 (planar, enamel), FMR56
An antenna coated with non-conductive material can be used if avoiding electrostatic charging
(e.g. through friction, cleaning, maintenance, strong medium flow).
Device type FMR51, FMR57 and Optional specification, ID Ox (Accessory Mounted) = OW
An antenna coated with non-conductive material can be used if avoiding electrostatic charging
(e.g. through friction, cleaning, maintenance, strong medium flow).
In the case of process connections made of polymeric material or with polymeric coatings, avoid electrostatic
charging of the plastic surfaces.
In the event of additional or alternative special varnishing on the housing or other metal parts:
Observe the danger of electrostatic charging and discharge.
Do not rub surfaces with a dry cloth.
11-13 Process
Connection
FMR50/56 UAE
XR0
XxG
Mounting bracket
Connection, without flange/mounting bracket
Slip on flange (different sizes)
FMR57 RxJ
XxJ
Thread, 316L
Align. device (different sizes)
*
1
FHX50 is approved according to DEK12.0046X or DEKRA 12ATEX0151X
ID Selected option Description
Ox Accessory
Mounted
FMR51/57 OW Horn protection, PTFE, no airpurge possible