Tankside Monitor NRF81 Operating menu
Endress+Hauser 135
User entry Signed floating-point number
Factory setting 0 mm
Additional information
Read access Operator
Write access Maintenance
Input value
Navigation Setup → Advanced setup → Input/output → Analog IP → Input value
Prerequisite Operating mode (→ 132) ≠ Disabled
Description Shows the value received via the analog input.
Additional information
Read access Operator
Write access -
Minimum probe temperature
Navigation Setup → Advanced setup → Input/output → Analog IP → Minimum probe
temperature
Prerequisite Operating mode (→ 132) = RTD temperature input
Description Minimum approved temperature of the connected probe. If the temperature falls below
this value, the W&M status will be 'invalid'.
User entry –213 to 927 °C
Factory setting –100 °C
Additional information
Read access Operator
Write access Maintenance
Maximum probe temperature
Navigation Setup → Advanced setup → Input/output → Analog IP → Maximum probe
temperature
Prerequisite Operating mode (→ 132) = RTD temperature input
Description Maximum approved temperature of the connected probe. If the temperature rises above
this value, the W&M status will be 'invalid'.