Status codes
The status codes comprise one byte and have got the following meaning:
If a status other than "GOOD" is sent to the device, the display indicates an error.
Status-
Code
Device status Significance Primary value Secondary value
0C Hex BAD device error X
0F Hex BAD device error X
1F Hex BAD out-of-service (target mode) X
40 Hex UNCERTAIN non-specific (simulation) X
47 Hex UNCERTAIN last usable value
(Fail-safe-Mode aktiv)
X
4B Hex UNCERTAIN Substitute set
(fail-Safe mode active)
X
4F Hex UNCERTAIN initial value
(fail-Safe mode active)
X
5C Hex UNCERTAIN Configuration error
(limits not set correctly)
X
80 Hex GOOD OK X X
84 Hex GOOD Active block alarm
(static revision counter incremented)
X
89 Hex GOOD LOW_LIM (alarm active) X
8A Hex GOOD HI_LIM (alarm active) X
8D Hex GOOD LOW_LOW_LIM (alarm active) X
8E Hex GOOD HI_HI_LIM (alarm active) X