Levelflex M FMP40 with PROFIBUS PA Commissioning
Endress + Hauser 81
6.8 Basic Setup with the ToF Tool
To carry out the basic setup with the ToF Tool operating program, proceed as follows:
• Start the ToF Tool operating program and establish a connection
• Select the "basic setup" function group in the navigation bar
The following display appears on the screen:
Basic Setup step 1/6:
•Status image
• Enter the measuring point description (TAG number).
L00-FMP4xxxx-20-00-00-en-001
!
Note!
• Each parameter that is changed must be confirmed with the RETURN key!
•The "Next" button moves you to the next screen display:
Basic Setup step 2/6:
• Enter the application parameters:
– tank properties (for a description, → ä 68)
– medium properties (for a description, → ä 69)
– process properties (for a description, → ä 70)