EasyManua.ls Logo

Endress+Hauser Proline Promag 300 - Page 18

Endress+Hauser Proline Promag 300
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XA01514D Proline Promag 300
18 Endress+Hauser
Intrinsically safe values
Order code for
"Output; input 1"
Output type Intrinsically safe values
"Output; input 1"
26 (+) 27 (–)
Option HA PROFIBUS PA Ex i Ex ic
U
i
= 32 V
l
i
= 570 mA
P
i
= 8.5 W
L
i
= 10 µH
C
i
= 5 nF
Option TA FOUNDATION
Fieldbus Ex i
Ex ic
U
i
= 32 V
l
i
= 570 mA
P
i
= 8.5 W
L
i
= 10 µH
C
i
= 5 nF
Order code for
"Output; input 2";
"Output; input 3"
Output type Intrinsically safe values
Output; input 2 Output; input 3
24 (+) 25 (–) 22 (+) 23 (–)
Option C Current output
4 to 20 mA Ex i
U
i
= 30 V
l
i
= 100 mA
P
i
= 1.25 W
L
i
= 0
C
i
= 0
Option G Pulse/frequency/
switch output Ex i
U
i
= 30 V
l
i
= 100 mA
P
i
= 1.25 W
L
i
= 0
C
i
= 0
Remote display DKX001
Basic specification, position 1,
2
Approval
Terminal assignment Basic specification, position 8
Display; Operation
Option O
Option BS 81, 82, 83, 84
U
n
= 3.3 V
I
n
= 150 mA

Other manuals for Endress+Hauser Proline Promag 300

Related product manuals