9.5 Defining the tag name
To enable quick identification of the measuring point within the system, you can enter a
unique designation using the Device tag parameter and thus change the factory setting.
Navigation path
Setup → Advanced setup → Device tag
9.5.1 Parameter overview with brief description
Parameter Prerequisite Description User entry Factory setting
Device tag The following option is selected
in the Header parameter:
Device tag
Enter the name for
the measuring
point.
Max. 32 characters,
such as letters,
numbers or special
characters (e.g. @,
%, /)
Prosonic Flow
9.6 Protecting settings from unauthorized access
The following options exist for protecting the configuration of the measuring device from
unintentional modification after commissioning:
• Write protection via access code (→ 37)
• Write protection via write protection switch (→ 38)
• Write protection via keypad lock (→ 32)
9.6.1 Write protection via access code
With the customer-specific access code, the parameters for the measuring device
configuration are write-protected and their values can no longer be changed via local
operation.
Navigation path
"Setup" menu → Advanced setup → Define access code
Structure of the submenu
Define access code
→
Define access code
Confirm access code
Defining the access code via local display
1. Navigate to the Define access code parameter.
2. Define a max. 4-digit numeric code as an access code.
3. Enter the access code again to confirm the code.
The -symbol appears in front of all write-protected parameters.