Proline t-mass 65F, 65I
10 Endress+Hauser
Supply voltage 85 to 260 V AC, 45 to 65 Hz
20 to 55 V AC, 45 to 65 Hz
16 to 62 V DC
Cable entries Power-supply and signal cables (inputs/outputs):
Cable entry M20 x 1.5, 8 to 12 mm (0.3 to 0.47")
Thread for cable entries, G 1/2", 1/2" NPT
Connecting cable for remote version:
Cable entry M20 x 1.5, 8 to 12 mm (0.3 to 0.47")
Thread for cable entries, G 1/2", 1/2" NPT
Remote version
cable specifications
2 x 2 x 0.5 mm² (AWG 20) PVC cable with common shield (2 twisted pairs)
Conductor resistance: 40 Ω/km [ 131.2 Ω /1000 ft]
Capacitance, core/shield: 0.001 μF/m [ 3.280 nF/ft]
Inductance: 0.9 μH/m [ 2.952 μH/ft]
•Operating voltage: 250 V
Temperature range: –40 to +105°C [–40 to +221°F]
Overall nominal diameter: 8.5 mm [0.335"]
Maximum cable length: 100 m [328 feet]
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010, the EMC
requirements of IEC/EN 61326, and NAMUR recommendation NE 21/43.
Note!
For cable specifications for hazardous area devices please see the additional Ex documentation.
Power consumption AC: 85 to 260 V = 18.2 W ; 20 to 55 V = 14 W ; (including sensor)
DC: 8 W (including sensor)
Power supply failure Lasting minimum 1 power cycle
EEPROM/HistoROM T-DAT saves measuring system data if the power supply fails
HistoROM S-DAT is on exchangeable data storage chip with sensor specific data (pipe type, nominal diam-
eter, serial number, flow conditioner, zero point, etc.)
Totalizer stops at the last value determined
Potential equalisation No measures necessary
For devices in hazardous areas please refer to the additional Ex documentation
Performance characteristics
Reference calibration
conditions
Accredited according to ISO/IEC 17025
Tracable to National Standards
Temperature controlled to within ±0.5°C (± 0.9°F) at atmospheric pressure and humidity controlled
Maximum measured error Flanged version:
±1.5 % of reading for 100 % to 20 % of full scale at reference conditions
±0.3 % of full scale for 20 % to 1 % of full scale at reference conditions
Insertion version:
±1.5 % of reading plus ±0.5 % of full scale
Note!
Calibration gas is normally air at reference calibration conditions and with a fully developed flow profile.
On site performance is dependant on the standard of installation.
Repeatability 0.5 % for velocities above 0.2 m/s (0.65 ft/s)
Response time Typically less than 2 seconds for 63 % of a given step change (in either direction).