Trouble-shooting RIA16
24 Endress+Hauser
11 Trouble-shooting
The following section provides you with an overview of possible causes of error to help you in
trouble-shooting.
11.1 Trouble-shooting instructions
#
Warning!
In the case of devices approved for use in explosion hazardous areas, error diagnosis cannot be
performed on the open device as the explosion protection would no longer be guaranteed.
11.2 Process error messages
Errors occurring during the self test or during operation are shown immediately on the display.
Acknowledgeable error messages are deleted after pressing a key. A fault is present if the hardware
for writing and reading data (EEPROM) is defective or if data cannot be read properly from the
EEPROM.
Error code Description
C561 Display overflow
F041 Sensor / cable open circuit (input < 3.6 mA or > 21 mA). Warning symbol ’Failure’ is displayed.
F045 Sensor error (2 mA < input 3,6 mA or input 21 mA). Warning symbol ’Failure’ is displayed.
F101 Signal underrange (input between 3.6 mA and 3.8 mA). Warning symbol ’Failure’ is displayed.
F102 Signal overrange (input between 20.5 mA and 21 mA). Warning symbol ’Failure’ is displayed.
F261 EEPROM error. Warning symbol ’Failure’ is displayed.
F282 Parameter data not successfully saved. Warning symbol ’Failure’ is displayed.
F283 Incorrect parameter data. Warning symbol ’Failure’ is displayed.
F431 Calibration values incorrect. Warning symbol ’Failure’ is displayed.