Hardware Installation and Maintenance Manual B1
Appendix B: Maintenance and
Troubleshooting
This section presents recommendations and solutions to common problems,
such as gas leaks, contamination, excessive sampling gas temperatures and
pressures, and electrical noise. Contact Service if your analyzer does not
appear to be hampered by one of these related problems, refer to “Service
Contact” on page B-31.
Gas Leaks
Probably the most common cause of erroneous measurements is outside air
leaking into the sample supply line. It is recommended the supply lines be
periodically leak tested, especially if the analyzer has been relocated or has
been replaced or returned to the factory for service and the supply lines have
been reconnected.
Contamination
Contamination and long exposure to high humidity are valid reasons for
periodically cleaning the gas sampling lines. Contamination in the gas sampling
lines can potentially find its way to the sample cell and deposit on the optics or
interfere with the measurement in some other way. Although the analyzer is
Class 3B invisible laser radiation when open. Avoid exposure to
the beam. Never open the sample cell unless directed to do so by
Service and the analyzer power is turned off.
The optical head has a seal and “WARNING” sticker to prevent
inadvertent tampering with the device. Do not attempt to
compromise the seal of the optical head assembly. Doing so will
result in loss of device sensitivity and inaccurate measurement
data. Repairs can then only be performed by the factory and are
not covered under warranty.
Do not use plastic tubing of any kind for sample lines. Plastic
tubing is permeable to moisture and other substances which can
contaminate the sample stream. SpectraSensors recommends
using 1/4 in. O.D x 0.035 in. wall thickness, seamless stainless
steel tubing.
Process samples may contain hazardous material in potentially
flammable and/or toxic concentrations. Personnel should have a
thorough knowledge and understanding of the physical properties
and safety precautions for the sample contents before operating
the SCS.

Table of Contents

Other manuals for Endress+Hauser SpectraSensors SS500

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser SpectraSensors SS500 and is the answer not in the manual?

Endress+Hauser SpectraSensors SS500 Specifications

General IconGeneral
BrandEndress+Hauser
ModelSpectraSensors SS500
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals