15
MN-37072 • Rev 2 • 05/16 • Designer Series Blanket Warmer
Cleaning and Preventative Maintenance
Monthly Checklist:
❏ Check the integrity of the door gasket. Are there any tears? Is the gasket
worn or loose? Ensure that the seal is tight to the body. Replace the gasket if
the integrity is compromised.*
❏ Check the air temperature sensor mounted in the interior of chamber. Is the
guard in place and is it fully secured to the appliance?
❏ Check the condition of the casters or feet. Are the components secure and
tightly threaded?
❏ Check the control panel overlay condition. Are there any tears or excessive
wear on the graphic? Does the control work properly when the buttons
are pushed?*
❏ Check the condition of the stacking hardware (if applicable), making sure
the mounting bolts and the hardware are secure.
Preventative Maintenance Checklist
Weekly Checklist:
❏ Inspect the condition of the plug and cord and replace if damaged.
❏ Remove the inserts and wash separately, set aside to dry before placing back
into the appliance.
❏ Blow dust from the interior, the outer vents and around the top of bonnet.
❏ Wipe down the interior.
❏ Check that the electronic control LEDs illuminate.*
❏ Check that the interior LED illuminates (if applicable).*
❏ Check the insert assembly. Check the blanket support assembly and the
shelf. Is the assembly in place? Are any pieces missing?*
Daily Checklist:
❏ Is the operation and care manual available?
❏ Has everyone been properly trained in the operation and safety instructions
of this appliance?
❏ Do not overload the appliance. Blankets must not exceed the height of the
support assembly. Allow 1" (25mm) gap between the top interior walls or
shelf and the blankets.
*Contact service for immediate repair.
Six-month Checklist:
❏ Is the set-point temperature comparable to the actual temperature
displayed? Check cavity air temperature with a quality thermocouple placed
1" (25mm) from the cavity sensor – not allowing it to touch any surface.
Monitor for approximately one hour in an empty interior.
Note: Blanket warmer temperature may fluctuate ±10° from set point.