16
MN-37072 • Rev 2 • 05/16 • Designer Series Blanket Warmer
Dimension Drawings
DC150 with standard feet
9.2
" (235mm)
IEC Cord Inlet
18.0" (456mm)
39.1" (994mm)
18.5" (470mm)
17.0" (431mm)
16.2" (412mm)
14.4" (366mm)
INSERT
10.1" (258mm)
CAVITY
9.6" (245mm)
INSERT
14.5" (368mm)
CAVITY
17.7" (450mm)
18.8" (477mm)
19.8" (504mm)
13.5
" (344mm)
IEC Cord Inlet
15.5" (394mm)
CAVITY
36.0" (914mm)
Cord length: 6 ft (1.83m) (est.)
22.7" (577mm)
Dimensions
(H x W x D)
With feet (standard): 17.0" x 18.5" x 22.7" (431mm x 470mm x 577mm)
With casters
(optional): 20.5" x 18.5" x 22.7" (520mm x 470mm x 577mm)
Capacity
1.5 
3
Weight**
(est.)
Net: 56 lbs (25 kg) Ship: 98 lbs (44 kg)
Electrical
*Other international plugs are available, contact factory for more information.
**Domestic ground shipping information. Contact factory for export weight and dimensions.
120 V.A.C. — 60 Hz, 1 ph
0.6 kW, 5.0 Amps
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
IPX-0
E471516
220 V.A.C. — 50Hz, 1 ph
0.6 kW, 2.7 Amps
Type B Equipment
CEE 7/7*
220-230V Plug
IPX-0
230 V.A.C. — 50 Hz, 1 ph
0.6 kW, 2.6 Amps
Type B Equipment
BS 1363 Plug*
(UK only)
CEE 7/7*
220-230V Plug
IPX-0
**Domestic ground shipping information. Contact factory for export weight and dimensions.
Clearance requirements:
3" (76mm) from rear
1" (25mm) from top and sides
3/4" (19mm) from bottom