Communication is established by clicking on “Read device”. The acknowledgement message
“Parameter read in” is displayed. If an error message is displayed instead, check the communication
parameters and cable connections between the PC and the Evaluation unit.
The edited data is transmitted to the Evaluation unit via “Program device”. Critical data concerning the
Modbus communication and the calibration must be conrmed before the parameters are transmitted
to the Evaluation unit:
If, when saving the the parameters in the Evaluation unit, the system calibration data is changed,
this action must be conrmed by checking “Overwrite calibration”.
If, when saving the the parameters in the Evaluation unit, the system interface parameters are
changed, this must be conrmed by checking the selection “Overwrite baud rate address”.
In addition, with the PC software,
the parameters of the Evaluation unit can be saved in a le ("Save conguration")
the parameters of the Evaluation unit can be loaded from a le ("Load conguration")
the parameters of the Evaluation unit can be printed via the set Windows standard printer (“Print
conguration”)
the measured values can be logged in a data logger le (enter the le name and storage rate, and
activate the data logger on the online display)
The software language can be set by right-clicking the “Sprache/Language/Langue” eld in the
bottom program line on “Deutsch/English/Français”.
Protection against unauthorised use:
The PC interface does not have a password prompt as it is assumed that only authorised personnel will
have access to the PC and the software. However, the password to operate the display can be read and
changed in menu 7. System under 7.6 Password.