1TEST PROGRAMS AND ADJUSTMENTS109 / 266
Art. No.: 80116-201
09 / 2004
Adjustment of Neurotest
board
Neurotest board
EE 30128-070
Activation test
Activation of the HF LED on the remote control takes place at intensity setting 0 using the rotary switch on
PC board EE 30121-442. At an intensity setting of 1 to 5 the NT LED is actuated by the CPU and the HF
LED must not light up. On the TUR version a check must be performed as to whether activation of the HF
is inhibited when intensity of 1 ... 5 is set. On the ZMK version, activation of the HF is possible even if
intensity 1 ... 5 has been set.