EasyManuals Logo

Erbe ICC 200 Service Manual

Erbe ICC 200
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #17 background imageLoading...
Page #17 background image
1TEST PROGRAMS AND ADJUSTMENTS17 / 266
Art. No.: 80116-201
09 / 2004
Test program no. 1
ICC 200, 300 only
General description
Using this test program, any front panel setting can be made. This setting is known as the basic setting.
After termination, the basic setting is stored internally. Only completely saved channels are stored. The
basic setting display flashes. Acknowledge by pressing any key.
Termination
Using “Power off”.
Switch off a channel
Using the “Down” key (8), set the lowest intensity. Once the lowest display is set, the seven-segment
display changes to “---”. This means that the channel has been switched off.
Basic setting and brief power failure
For a brief power failure of 15 seconds max., the front panel setting last set is displayed.
Activation of a channel is blocked if a channel has not been completely set or the basic setting was not
acknowledged.
The basic setting set at the factory corresponds to the FIXE basic setting.
FIXE basic setting
A FIXE basic setting is stored in the program. This basic setting is accepted and displayed if:
the stored front panel setting is invalid, e.g. through failure of the circuit memory or through battery
memory lost,
in Test program 1, the “All off” setting has been accepted.