1 TEST PROGRAMS AND ADJUSTMENTS26 / 266
Test program no. 7
Setting the NESSY version
General description
Using this program, you can set four NESSY versions. After starting the test program, you will see the
following displays when you activate key 3:
Display
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
NE.1 || or |
In the AUTO COAG field:
|| means a split neutral electrode,
| means of non-split neutral electrode.
Using the keys 8 (Up) or 9 (Down), you can set four different NESSY versions:
No.
AUTO CUT AUTO COAG 1 Remarks
1
NE.1 || or | for split and non-split electrodes
2
NE.2 | only for non-split electrodes
3NE.3 ||
only for split electrodes; acoustic (3 times) and
visual alarm
4NE.4 ||
only for split electrodes; acoustic (continuous
tone) and visual alarm
At the end of the test, the NESSY version number is stored. Version number 1 is the standard version set
when the unit is delivered. In case of memory failure, version no. 1 is automatically set. When switching on
the power, version no. 2, 3 or 4 is displayed briefly if this is set. The standard version no. 1 is not displayed.