1 TEST PROGRAMS AND ADJUSTMENTS22 / 266
Test program no. 5
Sound control for all available sounds
General description
Using this program, you can test and set all available sounds. After starting the program, you will see the
following display:
Display
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
ton 0 ... 4
Using the keys 8 (Up) or 9 (Down), you can set a test number from 0 to 4. The following correspondence
applies between the test numbers and the sounds:
Test no. Tones switched on Tone variants Remarks
0
Tone generation
switched off
1 Basic tone 1 Volume adjustable
2 Basic tone 2 Volume adjustable
3 Basic tone 3 Volume adjustable
4 Basic tone 4 Volume adjustable
Basic tone 1 to 3 Volume adjustable
Mixed tone: Basic tone 1 and 2 Volume adjustable
Mixed tone: Basic tone 1 and 3 Volume adjustable
Mixed tone: Basic tone 2 and 3 Volume adjustable
Basic tone 1 to 3 Alarm volume
Mixed tone: Basic tone 1 and 2 Alarm volume
Mixed tone: Basic tone 1 and 3 Alarm volume
Mixed tone: Basic tone 2 and 3 Alarm volume