1TEST PROGRAMS AND ADJUSTMENTS35 / 266
Art. No.: 80116-201
09 / 2004
Display program no. 12 (Start)
Display program no. 12:
Contact resistance at AUTO START, in standby operation.
Only valid for AUTO START 0, 1, 2 and in standby operation.
Display for ICC 350 and ICC 300:
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
uuu P2 P3 br
uuu Measurement value of the contact monitor voltage measurement
P2 Set power for AUTO COAG 1 [W]
P3 Set power for AUTO COAG 2 [W]
br Contact monitor
Display program no. 12:
Contact resistance for AUTO START, in active condition.
Only valid for AUTO START 0, 1, 2 and in active condition.
Display for ICC 350 and ICC 300:
AUTO CUT AUTO COAG 1 AUTO COAG 2 AUTO BIPOLAR
rdaP2P3br
rda Calculated value of the contact resistance [ohms]
P2 Set power for AUTO COAG 1 [W]
P3 Set power for AUTO COAG 2 [W]
br Contact monitor