7 Remedying malfunctions
30 / 46
80116-443
10.16
Display y.1 ECB fault Check IES 2 and VIO electrosurgi-
cal unit software versions for
compatibility.
Display y.2 Excess temperature (> 60 °C) Switch off the unit and wait until
the temperature in the interior of
the unit has returned to normal.
Display y.3 Insufficient temperature (< 0 °C) Switch off the unit and wait until
the temperature in the interior of
the unit has returned to normal.
Display y.4 Software error (check total) Reinstall software.
IMPORTANT! Software may only
be re-installed and updated by
Erbe Elektromedizin. Send unit for
repair.
Display y.5 Setup parameter error Restart unit.
Carry out re-initialisation.
Display y.6 Filter data erroneous Restart unit.
Carry out re-initialisation.
If necessary, replace filter.
Display y.7 Pressure sensor error Restart unit.
Replace control board.
Error Possible cause Action