39 / 46
9 Replacing components
80116-443
10.16
CHAPTER 9
Replacing components
Safety information
Replacing components
The "wearing parts", such as seals or similar components, are designed
such that, given normal operating conditions, it should not be necessary to
replace components in the first 10
years (except for filter and consum-
ables).
The following parts may have to be replaced during servicing:
Contact pins on the
contact circuit board
Remove main filter; do not open the housing! Remove the contact pins by
pulling them out firmly by hand.
Boards, fan and other internal
components
Having removed the housing cover, they can be removed with suitable
tools. When opening and closing the housing cover, be sure to be careful
of the grounded conductor fastening on the cover.
Software Software may only be updated by Erbe Elektromedizin using suitable pro-
gramming tools.
WARNING! Unplug the power cord from the outlet before exchanging parts of the
unit or cleaning it.
ATTENTION! Adjustments, technical tests, modifications, maintenance and repair
work may only be performed by Erbe or persons trained by Erbe. If the
work is not performed by trained persons, Erbe accepts no liability and
warranty rights become void.
ATTENTION! Only use original Erbe spare parts. The manufacturer accepts no liability
and the warranty rights becomes void if original spare parts are not
used.

Related product manuals