35 / 46
8 • Maintenance and servicing
80116-443
10.16
Activation with VIO electrosurgical unit (via the ECB)
Testing and measuring equipment
Test setup
• as with previous testing and also
• The power cord connects a VIO D or VIO S to the mains power supply.
• The IES 2 and the VIO are connected to one another via the ECB.
• Both units are switched on.
• The two-pedal footswitch with ReMode is connected to the VIO.
Test procedure
Activation via VIO 1. Perform the following settings on the VIO:
– Assign two-pedal footswitch to the bipolar receptacle.
– IES settings for bipolar receptacle: Cut suction: OFF, coag suction
80%, basic suction: OFF, run-on time 1 sec
2. Keep the COAG pedal of the VIO two-pedal footswitch pressed. The
suction must be activated.
3. Release the COAG pedal of the VIO two-pedal footswitch. The suction
must be deactivated.
4. Switch off VIO.
Erbe Art. No. Description
20322-101 IES 2 one pedal footswitch
20189-303 Dual-pedal footswitch with ReMode